Myra, from Australian company BioMolecular Systems (BMS), is an interesting entry into the liquid handling world on a number of axes. Liquid handling robots were once always bench-devouring monsters made by a small number of vendors whose names became synonymous with the technology. Later came OpenTrons, with the OT-2 loved by tinkerers but with a poor reputation among automation connoisseurs and the much better received Flex. The technology base keeps broadening, and there are easily a dozen U.S. or European manufacturers, plus a gaggle of Chinese ones. It's also the case that some of the major players are licensing out their technology, such as ONT building the now defunct ElysION using Tecan liquid handling technology.
BMS represents yet another new entrant, but with a very different physical profile than many liquid handlers. It's also aimed squarely as an affordable option for small throughput markets. The base Myra has only 6 deck positions and a $29K list price. There's also the Myra9, with nine deck positions. Accessories include integrating BMS's Mic qPCR (more below on this unusual device) as well as heating or cooling blocks. BMS will also custom 3D-print holders for the system. A mini HEPA filter and built-in UV LEDs should be appealing for users concerned about contamination. A four-channel pipette head is also an option.
Myra also comes with a camera and AI-powered software to take advantage of it. BMS claims that their system can detect bad deck layouts such as wrong labware in the wrong position. The camera and pressure-sensing in the pipettor are used for auto-calibration and verification of liquid transfers.
On the software side, they have a scheme ("MyraScript") enabling users to write Python code to run many details of the system, though this must be run inside their proprietary Workbench GUI interface. For remote controlling, there is a Remote Procedure Call (RPC) interface, though I haven't dug into that in detail.
The Mic unit is interesting. Rather than being plate-based, it is a spinning rotor holding 48 tubes. So no edge or corner effects! Rather than Peltier elements, induction heating and fans are used. Fixed magnets within the housing cause inductive currents and therefore heating in the rotor when the rotor is spun. Ramp rates of 4°C per second are achieved. Fans cool the rotor by forced convection, and both heating and cooling are enabled by the low thermal mass of the aluminum rotor. If integrated into a Myra, then the pipetting arm can load the tubes into the rotor from a deck position. Mic can be purchased with 2 or 4 optical channels.
Myra is definitely very different from the automation I am typically focused on; my last piece after all was exploring how to make instruments far more autonomous and there isn't an obvious path to Myra or Myra9. Sure, these have the precision to handle 384 well or even 1536 plates. But with only one or four pipetting channels, it's going to be slow for really large tasks.
After this paragraph I may need to open the doors on a Reconfigurable Automation Cart and whisper "peccavi!". There are, at least in the near future, going to be labs that won't be in the market for our autonomous laboratories, such as small pathology labs and small biotechs.
Plus, Myra might be inexpensive enough that you might let undergraduates near one. I never totally tanked an instrument as a student, but I did add dents to the inside of a clinical centrifuge. Oh, and put an invisible hole in a high vacuum line as a high school student. $30K ain't nothing, but perhaps in the range where higher level students might be allowed hands-on time.
Even in our lab, there's still the manner of preparing samples before they can go onto Nebula. We'd like to subsume all that, but it will take time. So there's still a need to go to the bench and make dilutions of stock solutions, or to reformat materials from non-SBS format containers such as Eppendorfs tubes, 15ml conicals, or 50ml conicals.
For my biochemical experiment, I needed to make five solutions to go on the system in three different plates. There's a tree of buffer preparation behind that. One required titration to the correct pH, which Myra doesn't appear to have a probe for. Also, many of the volumes would be too large. Definitely Myra could handle my enzyme and EDTA dilutions. Could I fill the three 384 well plates with Myra? Maybe, though the pipettors are limited to 50µl max, so for the 75µl fill that requires two passes. I hated doing that manually (thank you Liquidator for saving me from further suffering!) but perhaps it could go on Myra.
So maybe I'm just not the right audience. But with a low entry cost, compact design, and some advanced vision features, the Myra family could be very interesting in low to medium throughput labs running qPCR, short read sequencing, or other molecular biology protocols. Or for educational settings wanting a low-cost option to outfit a teaching lab.